NCT05995756: Evaluate the Safety and Efficacy of Sinocare CGM System Regarding Real Time Glucose Level Monitoring
From Jan 2022 to June 2022, 120 subjects (22-75 years of age, mean ±standard deviation \[SD\] of 53.20 ±14.49 years) with type 1 or type 2 diabetes participated in the study in P.R. China. Subjects wore two sensors in the abdomen which were paired with smart devices running a glucose monitoring mobile application (i3/H3 CGM APP). Subjects were asked to undergo one in-clinic visit (on one of the study days 2, 7-9, or 15) for 7-hour long YSI frequent sampling test (FST) periods, where intravenous (IV) blood samples were drawn every 15 min and analyzed using the YSI 2300 STAT Plus Glucose \& Lactate Analyzer (Yellow Springs Instrument, YSI Life Sciences, Yellow Springs, OH).
- Status
- COMPLETED
- Sponsor
- Sinocare
- Conditions
- Type 1 or Type 2 Diabetes
- Therapeutic areas
- Diabetes Mellitus, Type 2
- Enrollment
- 120
- Sites
- 3