NCT04380389: Accuracy of ScanWatch Pulse Oximeter With Profound Hypoxia

Performance testing of pulse oximeters. The protocol involves brief stable arterial oxygen desaturation in healthy volunteers and sampling arterial blood when a stable level of hypoxia has been attained. The blood sample is analyzed for oxygen saturation with a gold standard bench CO-oximeter. This protocol is aligned with the latest ISO and FDA guidance documents for pulse oximeter testing.

Status
COMPLETED
Sponsor
Withings
Conditions
Oximetry
Enrollment
14
Sites
1

Sites running this trial

Browse all clinical trials