Purpose: to evaluate the immediate effect of automated oxygen titration compared to usual fixed-dose oxygen treatment during exercise in patients with COPD on long-term oxygen treatment. Methods: The study will be conducted as a double blinded randomized crossover trial with two arms. 40 ambulatory patients with COPD and home oxygen treatment will be included from AHH Hospital's catchment area. The patients will conduct two Endurance Shuttle Walk Tests (ESWT) in a crossover design using an O2matic device to deliver a variable oxygen dosage set at an SpO2-target of 90-94% and an O2-flow of 0 - 15 liters/min and using the patients´ usual fixed-dose oxygen delivery, in a randomized order. In both arms O2matic will monitor pulse rate and SpO2 continuously during the test, but only in the automated oxygen titration arm will O2matic adjust oxygen flow. The patient and the physiotherapist supervising the tests will be blinded to the oxygen dose. Primary outcome is the changes in perceived dyspnea intensity using Borg CR10 scale between walking with automated titration compared to fixed-dose treatment. Secondary outcomes are differences in walking time, the average oxygen consumption between automated oxygen titration and fixed-dose treatment and difference in time spent within acceptable SpO2-interval.
- Status
- COMPLETED
- Sponsor
- Hvidovre University Hospital
- Conditions
- Chronic Obstructive Pulmonary Disease, Respiratory Failure
- Therapeutic areas
- Dyspnea, Pulmonary Disease, Chronic Obstructive, Respiratory Insufficiency
- Enrollment
- 35
- Sites
- 1