NCT03256955: Validation of the TOF Cuff Monitor® Which Measures Neuromuscular Block on the Upper Arm
Neuromuscular blocking agents (NMBAs) are frequently used in anesthesia and quantitative neuromuscular monitoring is standard care. The TOF WATCH SX® monitor is considered as one of the reference monitoring devices in clinical research and clinical practice. With this monitor the ulnar nerve is stimulated at the wrist and the force of the movement of the thumb is measured with acceleromyography. This method requires freedom of movement of the patient's thumb. Unfortunately this is not always possible due to the constraints of patient positioning during the operation. The TOF Cuff® monitor is a modified non-invasive blood pressure cuff that incorporates stimulating electrodes in its inner surface and is based on the stimulation of the peripheral nerve in the arm (brachial plexus, ulnar and median nerves principally). The evoked neuromuscular activity is recorded through the changes in pressure generated in the inner part of the cuff by the muscular activity after the stimulus. Moreover, this device can be used for non-invasive reading of the blood pressure. This device has been validated with mechanomyography, but was never been compared with acceleromyography, which is the most common used neuromuscular monitoring method.
- Status
- COMPLETED
- Sponsor
- Christoph Czarnetzki
- Conditions
- Neuromuscular Blockade
- Enrollment
- 40
- Sites
- 1