NCT02094859: GlucoClearTM System ICU Study
The primary purpose of this Study is evaluation of the safety and performance of the GlucoClear System (System). The primary safety objective will be assessed by: • Evaluation of any serious adverse device effects upon Sensor insertion through Sensor removal and a follow-up assessment one week (7 ± 3 days) after Sensor removal. The primary performance objective will be evaluated using the following criteria: 15/20% Performance: Blood glucose measurements from the System must meet the ISO 15197 15/20% accuracy criteria (at least 95% of the matched pairs within ± 15 mg/dL of Comparator values ≤ 75 mg/dL and within ±20% of Comparator values \> 75 mg/dL).
- Status
- COMPLETED
- Sponsor
- Edwards Lifesciences
- Conditions
- Accurate Blood Glucose Monitoring in Critical Care Setting
- Enrollment
- 50
- Sites
- 2