NCT00933738: Evaluation of the Accuracy of the INRatio Prothrombin Time (PT) Monitoring System With a New Test Strip for the Oral Anticoagulation Therapy Patient in the Presence of Heparin and Low Molecular Weight Heparin (e.g., Enoxaparin or Dalteparin)
This is a multi-center study designed to evaluate the heparin insensitivity of the INRatio Prothrombin Time (PT) Monitoring System, utilizing an INRatio test strip additionally modified for low sample volume. The INRatio test strip is used for the quantitative determination of PT and International Normalized Ratio (INR) results in fingerstick blood from subjects on oral anticoagulation therapy (OAT) with warfarin. This study is designed to evaluate the accuracy of the modified INRatio test strip during heparin-warfarin bridge therapy with unfractionated heparin (UH) or low molecular weight heparin (LMWH), such as enoxaparin or dalteparin. These INR results will be compared to the INR results obtained on plasma from the same subjects as analyzed at a central laboratory with the heparin-insensitive reference method: the Sysmex CA-560 System. The levels of UH or LMWH in the plasma samples will be assessed using activated partial thromboplastin time (aPTT) and anti-factor-Xa assays respectively.
- Status
- COMPLETED
- Sponsor
- Abbott RDx Cardiometabolic
- Conditions
- Coagulation
- Therapeutic areas
- Thrombosis
- Enrollment
- 150
- Sites
- 13
Sites running this trial
- University of Colorado Hospital, Aurora, CO
- Henry Ford Hospital, Detroit, MI
- University of New Mexico Hospital, Albuquerque, NM
- Swedish Medical Center, Seattle, WA
- University of Colorado at Denver, Aurora, CO
- Wenatchee Valley Medical Center, Wenatchee, WA
- New York Methodist Hospital, Brooklyn, NY
- Michigan Heart And Vascular Specialists, Petoskey, MI
- Northwest Houston Heart Center, Tomball, TX
- International Heart Institute of Montana, Missoula, MT
- John T Mather Memorial Hospital, Port Jefferson, NY
- Emergency Medicine Research Group, Lansing, MI